Recall Z-1265-2016— Biomet, Inc.
Class II · Terminated
Class II FDA medical device recall by Biomet, Inc., initiated 2016-02-05, terminated 2017-03-28. It names one FDA 510(k) clearance: K141407. Product: Vanguard XP Knee System Porous Femoral Component various sizes and orientation a total knee replacement system. prosthesis, knee, patella/femorotibial, semi-constrained, uncemented, porous, coated, polymer/metal/polymer. Reason: The anterior/posterior (AP) dimension may be oversized by 0.016. If the femoral component does not fit as tightly as intended, bony ingrowth may not occur causing loosening of the femoral component. A revision surgery may be necessary..
- Recalling firm
- Biomet, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-02-05
- Terminated
- 2017-03-28
- Firm location
- Warsaw, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Vanguard XP Knee System Porous Femoral Component various sizes and orientation a total knee replacement system. prosthesis, knee, patella/femorotibial, semi-constrained, uncemented, porous, coated, polymer/metal/polymer
Reason for recall
The anterior/posterior (AP) dimension may be oversized by 0.016. If the femoral component does not fit as tightly as intended, bony ingrowth may not occur causing loosening of the femoral component. A revision surgery may be necessary.