Recall Z-1264-2022— Olympus Corporation of the Americas
Class II · Ongoing
Class II FDA medical device recall by Olympus Corporation of the Americas, initiated 2022-04-20. It names one FDA 510(k) clearance: K172298. Product: Olympus URETERO-RENO FIBERSCOPE Model: URF-P6R. Reason: Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis..
- Recalling firm
- Olympus Corporation of the Americas
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-04-20
- Firm location
- Center Valley, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Olympus URETERO-RENO FIBERSCOPE Model: URF-P6R
Reason for recall
Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.