Recall Z-1262-2014— Zimmer, Inc.

Class II · Terminated

Class II FDA medical device recall by Zimmer, Inc., initiated 2014-02-25, terminated 2015-03-24. It names one FDA 510(k) clearance: K936159. Product: Natural-Knee¿ II System: POLY ULTRA TIB LT SZ00/9M POLY ULTRA TIB LT SZ0/9MM POLY ULTRA TIB LT SZ0/19M POLY ULTRA TIB RT SZ0/13M POLY ULTRA TIB RT SZ0/16M ALL POLY ULTRA TIB LT SZ1 ALL POLY ULTRA TIB RT SZ1 ALL POLY ULTRA TIB LT SZ2 ALL POLY ULTRA TIB LT SZ2 ALL POLY ULTRA TIB RT SZ2 ALL POLY ULTRA TIB LT SZ3 ALL POLY ULTRA TIB RT SZ3 ALL POLY ULTRA TIB RT SZ5. Reason: A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device..

Recalling firm
Zimmer, Inc.
Classification
Class II
Status
Terminated
Initiated
2014-02-25
Terminated
2015-03-24
Firm location
Warsaw, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Natural-Knee¿ II System: POLY ULTRA TIB LT SZ00/9M POLY ULTRA TIB LT SZ0/9MM POLY ULTRA TIB LT SZ0/19M POLY ULTRA TIB RT SZ0/13M POLY ULTRA TIB RT SZ0/16M ALL POLY ULTRA TIB LT SZ1 ALL POLY ULTRA TIB RT SZ1 ALL POLY ULTRA TIB LT SZ2 ALL POLY ULTRA TIB LT SZ2 ALL POLY ULTRA TIB RT SZ2 ALL POLY ULTRA TIB LT SZ3 ALL POLY ULTRA TIB RT SZ3 ALL POLY ULTRA TIB RT SZ5

Reason for recall

A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.