Recall Z-1257-2020— Ormco/Sybronendo

Class II · Terminated

Class II FDA medical device recall by Ormco/Sybronendo, initiated 2020-01-17, terminated 2024-09-30. FDA's recall record names no specific 510(k) clearance for this event. Product: SnapLink, Ref Part No. 438-2190, L6R SnapLink, Slot .022, TQ--28, Ang +2, Rot +2, Qty 10, Made in Mexico, CE 0086, Rx Only. Reason: The action is being taken due to an increase in complaints relating to the lower buccal tube slide doors not staying close. In the event of functional failure, broken/damage buccal tubes it may cause the tooth to rotate, possibly delay treatment and inconvenience the patient and clinician. If not noticed it may extend treatment by approximately 1 -3 months..

Recalling firm
Ormco/Sybronendo
Classification
Class II
Status
Terminated
Initiated
2020-01-17
Terminated
2024-09-30
Firm location
Glendora, CA, United States

Product description

SnapLink, Ref Part No. 438-2190, L6R SnapLink, Slot .022, TQ--28, Ang +2, Rot +2, Qty 10, Made in Mexico, CE 0086, Rx Only

Reason for recall

The action is being taken due to an increase in complaints relating to the lower buccal tube slide doors not staying close. In the event of functional failure, broken/damage buccal tubes it may cause the tooth to rotate, possibly delay treatment and inconvenience the patient and clinician. If not noticed it may extend treatment by approximately 1 -3 months.