Recall Z-1257-2016— ICU Medical, Inc.

Class II · Terminated

Class II FDA medical device recall by ICU Medical, Inc., initiated 2016-02-25, terminated 2016-08-26. It names one FDA 510(k) clearance: K131549. Product: 110" (279 cm) Appx 13.8 ml, 20 Drop Admin Set w/ Integrated ChemoLock Port Drip Chamber, Clave, Rotating Luer, Bag Hanger, Item No. AHCL3300 sterile devices intended for the infusion and withdrawal of fluids.. Reason: ICU Medical has identified an issue with a connection in an isolated number of I.V. infusion administration sets that incorporate the ChemLock additive port. In these sets, the connection between the Drip Chamber or Bag Spike and the ChemoLock port has the potential for developing a leak..

Recalling firm
ICU Medical, Inc.
Classification
Class II
Status
Terminated
Initiated
2016-02-25
Terminated
2016-08-26
Firm location
San Clemente, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

110" (279 cm) Appx 13.8 ml, 20 Drop Admin Set w/ Integrated ChemoLock Port Drip Chamber, Clave, Rotating Luer, Bag Hanger, Item No. AHCL3300 sterile devices intended for the infusion and withdrawal of fluids.

Reason for recall

ICU Medical has identified an issue with a connection in an isolated number of I.V. infusion administration sets that incorporate the ChemLock additive port. In these sets, the connection between the Drip Chamber or Bag Spike and the ChemoLock port has the potential for developing a leak.

Data sourced from openFDA. This site is unofficial and independent of the FDA.