Recall Z-1256-2022— Steris Corporation

Class II · Terminated

Class II FDA medical device recall by Steris Corporation, initiated 2022-04-22, terminated 2024-11-04. FDA's recall record names no specific 510(k) clearance for this event. Product: Reliance Synergy Washer/Disinfector. Reason: The firm identified that when the electrical contactor component in the drying chamber of the Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire..

Recalling firm
Steris Corporation
Classification
Class II
Status
Terminated
Initiated
2022-04-22
Terminated
2024-11-04
Firm location
Mentor, OH, United States

Product description

Reliance Synergy Washer/Disinfector

Reason for recall

The firm identified that when the electrical contactor component in the drying chamber of the Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire.

Data sourced from openFDA. This site is unofficial and independent of the FDA.