Recall Z-1254-2021— Mindray DS USA, Inc. dba Mindray North America
Class II · Ongoing
Class II FDA medical device recall by Mindray DS USA, Inc. dba Mindray North America, initiated 2021-02-22. It names one FDA 510(k) clearance: K193391. Product: Mindray BeneVision Distributed Monitoring System( DMS) 1. p/n 115-051209-00 2. p/n 115-034414-00 and 115-034413-00 when used with the following software versions: 03.04.00 04.01.00 04.03.00 04.04.00 04.04.01. Reason: BeneVision DMS may intermittently freeze and require a manual reboot after which normal operation resumes. If a freeze occur, patients monitored on a telemetry transmitter will no longer communicate data to the BeneVision DMS.
- Recalling firm
- Mindray DS USA, Inc. dba Mindray North America
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2021-02-22
- Firm location
- Mahwah, NJ, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Mindray BeneVision Distributed Monitoring System( DMS) 1. p/n 115-051209-00 2. p/n 115-034414-00 and 115-034413-00 when used with the following software versions: 03.04.00 04.01.00 04.03.00 04.04.00 04.04.01
Reason for recall
BeneVision DMS may intermittently freeze and require a manual reboot after which normal operation resumes. If a freeze occur, patients monitored on a telemetry transmitter will no longer communicate data to the BeneVision DMS
Data sourced from openFDA. This site is unofficial and independent of the FDA.