Recall Z-1254-2017— Siemens Healthcare Diagnostics, Inc

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc, initiated 2017-01-10, terminated 2019-07-16. It names one FDA 510(k) clearance: K961510. Product: Siemens ADVIA Centaur Systems Calibrator A Kits, Cat No. 04800646, SMN 10285903 (2 pack)Cat No. 04800735, SMN 10285904 (6 pack) Cat No. 04800840, SMN 10285905 (6 pack - REF). Reason: Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur FT4 when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT Systems. In addition, Siemens Healthcare Diagnostics confirmed the potential for calibration failures due to above limit calibrator RLU %CVs when using Calibrator A kit lots ending in 90 with the FT4 assay. The performance of the FT4 assay when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur CP System is not affected. Customers may continue to use Calibrator A kit lots ending in 90 for the ADVIA Centaur Systems FT3, T3, T4 and TUp assays..

Recalling firm
Siemens Healthcare Diagnostics, Inc
Classification
Class II
Status
Terminated
Initiated
2017-01-10
Terminated
2019-07-16
Firm location
East Walpole, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Siemens ADVIA Centaur Systems Calibrator A Kits, Cat No. 04800646, SMN 10285903 (2 pack)Cat No. 04800735, SMN 10285904 (6 pack) Cat No. 04800840, SMN 10285905 (6 pack - REF)

Reason for recall

Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur FT4 when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT Systems. In addition, Siemens Healthcare Diagnostics confirmed the potential for calibration failures due to above limit calibrator RLU %CVs when using Calibrator A kit lots ending in 90 with the FT4 assay. The performance of the FT4 assay when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur CP System is not affected. Customers may continue to use Calibrator A kit lots ending in 90 for the ADVIA Centaur Systems FT3, T3, T4 and TUp assays.