Recall Z-1254-2014— Zimmer, Inc.
Class II · Terminated
Class II FDA medical device recall by Zimmer, Inc., initiated 2014-02-25, terminated 2015-03-24. It names one FDA 510(k) clearance: K931651. Product: Natural-Knee¿ Congruent Tibial Insert with Screw: SF-CON TIB INS LT SZ1/2 1 SF-CON TIB INS RT SZ1/2 9 SF-CON TIB INS RT SZ3/4/5. Reason: A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device..
- Recalling firm
- Zimmer, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-02-25
- Terminated
- 2015-03-24
- Firm location
- Warsaw, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Natural-Knee¿ Congruent Tibial Insert with Screw: SF-CON TIB INS LT SZ1/2 1 SF-CON TIB INS RT SZ1/2 9 SF-CON TIB INS RT SZ3/4/5
Reason for recall
A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
Data sourced from openFDA. This site is unofficial and independent of the FDA.