Recall Z-1251-2023— BioFire Diagnostics, LLC
Class II · Ongoing
Class II FDA medical device recall by BioFire Diagnostics, LLC, initiated 2023-02-01. It names one FDA 510(k) clearance: K193519. Product: In Vitro Diagnostic (IVD) Panel BioFire BCID2 Panel - REF RFIT-ASY-0147. Reason: Due to temperature excursions with products (products being stored outside of the required temperature specifications) that may cause a reduction in performance. The temperature excursion was as a result of shipping delays..
- Recalling firm
- BioFire Diagnostics, LLC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-02-01
- Firm location
- Salt Lake City, UT, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
In Vitro Diagnostic (IVD) Panel BioFire BCID2 Panel - REF RFIT-ASY-0147
Reason for recall
Due to temperature excursions with products (products being stored outside of the required temperature specifications) that may cause a reduction in performance. The temperature excursion was as a result of shipping delays.
Data sourced from openFDA. This site is unofficial and independent of the FDA.