Recall Z-1249-2022— Getinge Usa Sales Inc
Class II · Ongoing
Class II FDA medical device recall by Getinge Usa Sales Inc, initiated 2022-04-13. FDA's recall record names no specific 510(k) clearance for this event. Product: Getinge CM320 WUWD Series Washer Disinfector-For use as a multi-chamber washer-disinfector for the washing, disinfecting and drying surgical instruments, Model: CM320 WUWD. Reason: Verification of installation not documented or may be incomplete, could result in harm to the user such as skin burns, smoke inhalation, electric shock, or a crush or pinch injury or slip and fall injury. May lead to a patient procedural delay and/or infection.
- Recalling firm
- Getinge Usa Sales Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-04-13
- Firm location
- Wayne, NJ, United States
Product description
Getinge CM320 WUWD Series Washer Disinfector-For use as a multi-chamber washer-disinfector for the washing, disinfecting and drying surgical instruments, Model: CM320 WUWD
Reason for recall
Verification of installation not documented or may be incomplete, could result in harm to the user such as skin burns, smoke inhalation, electric shock, or a crush or pinch injury or slip and fall injury. May lead to a patient procedural delay and/or infection
Data sourced from openFDA. This site is unofficial and independent of the FDA.