Recall Z-1248-2016— Maquet Datascope Corp - Cardiac Assist Division
Class II · Terminated
Class II FDA medical device recall by Maquet Datascope Corp - Cardiac Assist Division, initiated 2015-12-16, terminated 2018-06-01. It names one FDA 510(k) clearance: K112372. Product: Cardiosave Hybrid and rescue IABP. Reason: Maquet has recieved information that in some Cardiosave IABPS, the scroll compressor did not meet the specifications for output pressure or vacuum at specific flow rates..
- Recalling firm
- Maquet Datascope Corp - Cardiac Assist Division
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-12-16
- Terminated
- 2018-06-01
- Firm location
- Mahwah, NJ, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Cardiosave Hybrid and rescue IABP
Reason for recall
Maquet has recieved information that in some Cardiosave IABPS, the scroll compressor did not meet the specifications for output pressure or vacuum at specific flow rates.
Data sourced from openFDA. This site is unofficial and independent of the FDA.