Recall Z-1248-2015— GE Healthcare
Class II · Terminated
Class II FDA medical device recall by GE Healthcare, initiated 2015-01-30, terminated 2015-09-15. It names one FDA 510(k) clearance: K993090. Product: GE Healthcare Proteus XR/a with wall stand model number 2260354. Intended for use in generating radiographic images of human anatomy in all general purpose diagnostic procedures. This device is not intended for mammographic applications.. Reason: Potential breakage of the steel cable which supports the bucky device in certain wall stands of Proteus XR/a X-ray imaging systems. A fall of a bucky while the system is in use could result in an injury to a patient or operator..
- Recalling firm
- GE Healthcare
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-01-30
- Terminated
- 2015-09-15
- Firm location
- Waukesha, WI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
GE Healthcare Proteus XR/a with wall stand model number 2260354. Intended for use in generating radiographic images of human anatomy in all general purpose diagnostic procedures. This device is not intended for mammographic applications.
Reason for recall
Potential breakage of the steel cable which supports the bucky device in certain wall stands of Proteus XR/a X-ray imaging systems. A fall of a bucky while the system is in use could result in an injury to a patient or operator.