Recall Z-1247-2026— Vortex Surgical Inc.
Class II · Ongoing
Class II FDA medical device recall by Vortex Surgical Inc., initiated 2025-12-16. FDA's recall record names no specific 510(k) clearance for this event. Product: Vortex Disposable Forceps and Cannula 1. 23GA ACTU8 Forceps Adaptive VS0740.23 2. 25GA ACTU8 Forceps Adaptive VS0740.25 3. 25GA ACTU8 Forceps ILM VS0744.25 4. 25GA ACTU8 Forceps ILM -Myopic VS0744.25-35 5. 25GA ACTU8 Forceps Pro-Grip VS0713.25 6. 25GA ACTU8 Talon Forceps ILM VS0752.25 7. 25GA ACTU8 Talon Forceps Pro-Grip VS0751.25 8. 27GA ACTU8 Forceps ILM VS0744.27 9. 23GA Soft Tip Cannula VS0200.23 10. 23GA Soft Tip Cannula - 0.5mm VS0205.23 11. 25GA Soft Tip Cannula VS0200.25 12. 25GA Soft Tip Cannula - 0.5mm VS0205.25 13. 25GA Soft Tip Cannula - 0.5mm - Bulk VS0205.25B 14. 25GA Soft Tip Cannula-Bulk VS0200.25B 15. 27GA Soft Tip Cannula VS0200.27. Reason: There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection..
- Recalling firm
- Vortex Surgical Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-12-16
- Firm location
- Saint Charles, MO, United States
Product description
Vortex Disposable Forceps and Cannula 1. 23GA ACTU8 Forceps Adaptive VS0740.23 2. 25GA ACTU8 Forceps Adaptive VS0740.25 3. 25GA ACTU8 Forceps ILM VS0744.25 4. 25GA ACTU8 Forceps ILM -Myopic VS0744.25-35 5. 25GA ACTU8 Forceps Pro-Grip VS0713.25 6. 25GA ACTU8 Talon Forceps ILM VS0752.25 7. 25GA ACTU8 Talon Forceps Pro-Grip VS0751.25 8. 27GA ACTU8 Forceps ILM VS0744.27 9. 23GA Soft Tip Cannula VS0200.23 10. 23GA Soft Tip Cannula - 0.5mm VS0205.23 11. 25GA Soft Tip Cannula VS0200.25 12. 25GA Soft Tip Cannula - 0.5mm VS0205.25 13. 25GA Soft Tip Cannula - 0.5mm - Bulk VS0205.25B 14. 25GA Soft Tip Cannula-Bulk VS0200.25B 15. 27GA Soft Tip Cannula VS0200.27
Reason for recall
There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
Data sourced from openFDA. This site is unofficial and independent of the FDA.