Recall Z-1245-2025— Micro Therapeutics, Inc.
Class I · Ongoing
Class I FDA medical device recall by Micro Therapeutics, Inc., initiated 2025-01-30. FDA's recall record names no specific 510(k) clearance for this event. Product: Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3-027-350-14, PED3-027-350-16, PED3-027-350-20, PED3-027-350-25, PED3-027-400-12, PED3-027-400-14, PED3-027-400-16, PED3-027-400-18, PED3-027-400-20, PED3-027-400-25, PED3-027-400-30, PED3-027-450-12, PED3-027-450-14, PED3-027-450-16, PED3-027-450-18, PED3-027-450-20, PED3-027-450-25, PED3-027-450-30, PED3-027-450-40, PED3-027-500-14, PED3-027-500-16, PED3-027-500-18, PED3-027-500-20, PED3-027-500-25, PED3-027-500-30, PED3-027-500-40, PED3-027-550-16, PED3-027-550-18, PED3-027-550-20, PED3-027-550-30, PED3-027-550-40, PED3-027-550-50, PED3-027-600-16, PED3-027-600-18, PED3-027-600-20, PED3-027-600-30, PED3-027-600-40, PED3-027-600-50. Reason: Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age.
- Recalling firm
- Micro Therapeutics, Inc.
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2025-01-30
- Firm location
- Irvine, CA, United States
Product description
Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3-027-350-14, PED3-027-350-16, PED3-027-350-20, PED3-027-350-25, PED3-027-400-12, PED3-027-400-14, PED3-027-400-16, PED3-027-400-18, PED3-027-400-20, PED3-027-400-25, PED3-027-400-30, PED3-027-450-12, PED3-027-450-14, PED3-027-450-16, PED3-027-450-18, PED3-027-450-20, PED3-027-450-25, PED3-027-450-30, PED3-027-450-40, PED3-027-500-14, PED3-027-500-16, PED3-027-500-18, PED3-027-500-20, PED3-027-500-25, PED3-027-500-30, PED3-027-500-40, PED3-027-550-16, PED3-027-550-18, PED3-027-550-20, PED3-027-550-30, PED3-027-550-40, PED3-027-550-50, PED3-027-600-16, PED3-027-600-18, PED3-027-600-20, PED3-027-600-30, PED3-027-600-40, PED3-027-600-50
Reason for recall
Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age
Data sourced from openFDA. This site is unofficial and independent of the FDA.