Recall Z-1243-2025— GE Healthcare (China) Co., Ltd.

Class II · Ongoing

Class II FDA medical device recall by GE Healthcare (China) Co., Ltd., initiated 2025-01-31. FDA's recall record names no specific 510(k) clearance for this event. Product: SIGNA MR380, NMRI system, Non-US Only. Reason: For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant..

Recalling firm
GE Healthcare (China) Co., Ltd.
Classification
Class II
Status
Ongoing
Initiated
2025-01-31
Firm location
Beijing, N/A, China

Product description

SIGNA MR380, NMRI system, Non-US Only

Reason for recall

For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.