Recall Z-1240-2016— Siemens Healthcare Diagnostics, Inc

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc, initiated 2016-01-13, terminated 2018-01-08. FDA's recall record names no specific 510(k) clearance for this event. Product: Siemens ADVIA Centaur Enhance Estradiol Test Code eE2 Catalog Number: 10490889, 10491445, 10697757 Product Usage: For in vitro diagnostic use in the quantitative determination of estradiol in human serum and plasma (heparinized and EDTA) using the ADVIA Centaur¿, ADVIA Centaur XP, ADVIA Centaur XPT and ADVIA Centaur CP systems.. Reason: The drug fulvestrant (Faslodex¿) may cause falsely elevated estradiol results..

Recalling firm
Siemens Healthcare Diagnostics, Inc
Classification
Class II
Status
Terminated
Initiated
2016-01-13
Terminated
2018-01-08
Firm location
East Walpole, MA, United States

Product description

Siemens ADVIA Centaur Enhance Estradiol Test Code eE2 Catalog Number: 10490889, 10491445, 10697757 Product Usage: For in vitro diagnostic use in the quantitative determination of estradiol in human serum and plasma (heparinized and EDTA) using the ADVIA Centaur¿, ADVIA Centaur XP, ADVIA Centaur XPT and ADVIA Centaur CP systems.

Reason for recall

The drug fulvestrant (Faslodex¿) may cause falsely elevated estradiol results.

Data sourced from openFDA. This site is unofficial and independent of the FDA.