Recall Z-1233-2016— Roche Diagnostics Operations, Inc.

Class II · Terminated

Class II FDA medical device recall by Roche Diagnostics Operations, Inc., initiated 2016-02-23, terminated 2018-01-11. FDA's recall record names no specific 510(k) clearance for this event. Product: cobas p 512 pre-analytical system Pre-analytical sample handling that includes de-capping, aliquoting and sorting of samples for analysis.. Reason: Due to a false triggering or detection of the lifting gripper READY signal, sample tubes are not correctly placed back in the Rack Tube Transport (RTT) after the decapping process. Therefore, open sample tubes can be dropped in the cobas p 512, spilling the sample material..

Recalling firm
Roche Diagnostics Operations, Inc.
Classification
Class II
Status
Terminated
Initiated
2016-02-23
Terminated
2018-01-11
Firm location
Indianapolis, IN, United States

Product description

cobas p 512 pre-analytical system Pre-analytical sample handling that includes de-capping, aliquoting and sorting of samples for analysis.

Reason for recall

Due to a false triggering or detection of the lifting gripper READY signal, sample tubes are not correctly placed back in the Rack Tube Transport (RTT) after the decapping process. Therefore, open sample tubes can be dropped in the cobas p 512, spilling the sample material.