Recall Z-1232-2023— Datascope Corp.
Class II · Ongoing
Class II FDA medical device recall by Datascope Corp., initiated 2023-02-07. It names one FDA 510(k) clearance: K181122. Product: Cardiosave Rescue. Model Number: 0998-00-0800-83.. Reason: Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted..
- Recalling firm
- Datascope Corp.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-02-07
- Firm location
- Mahwah, NJ, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Cardiosave Rescue. Model Number: 0998-00-0800-83.
Reason for recall
Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.
Data sourced from openFDA. This site is unofficial and independent of the FDA.