Recall Z-1231-2023— Datascope Corp.
Class II · Ongoing
Class II FDA medical device recall by Datascope Corp., initiated 2023-02-07. It names one FDA 510(k) clearance: K181122. Product: Cardiosave Hybrid Intra-Aortic Balloon Pump. Model Numbers: D998-00-0800-31, D998-00-0800-32, D998-00-0800-33, D998-00-0800-35, D998-00-0800-36, D998-00-0800-45, D998-00-0800-52, D998-00-0800-53, D998-00-0800-55, D998-00-0800-65, D998-00-0800-83, D998-UC-0800-31, D998-UC-0800-33, D998-UC-0800-52, D998-UC-0800-53, D998-UC-0800-55.. Reason: Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted..
- Recalling firm
- Datascope Corp.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-02-07
- Firm location
- Mahwah, NJ, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Cardiosave Hybrid Intra-Aortic Balloon Pump. Model Numbers: D998-00-0800-31, D998-00-0800-32, D998-00-0800-33, D998-00-0800-35, D998-00-0800-36, D998-00-0800-45, D998-00-0800-52, D998-00-0800-53, D998-00-0800-55, D998-00-0800-65, D998-00-0800-83, D998-UC-0800-31, D998-UC-0800-33, D998-UC-0800-52, D998-UC-0800-53, D998-UC-0800-55.
Reason for recall
Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.