Recall Z-1229-2022— GE Healthcare, LLC

Class II · Ongoing

Class II FDA medical device recall by GE Healthcare, LLC, initiated 2022-04-28. It names one FDA 510(k) clearance: K131977. Product: Centricity Universal Viewer Zero Footprint Client. Reason: Potential for Distance and Area measurements to display inaccurate measurement values when performed on magnified images, and specific to ZFP, Distance and Area measurements can display inaccurate measurement values when performed on lossy images that are scaled down from their original resolution..

Recalling firm
GE Healthcare, LLC
Classification
Class II
Status
Ongoing
Initiated
2022-04-28
Firm location
Waukesha, WI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Centricity Universal Viewer Zero Footprint Client

Reason for recall

Potential for Distance and Area measurements to display inaccurate measurement values when performed on magnified images, and specific to ZFP, Distance and Area measurements can display inaccurate measurement values when performed on lossy images that are scaled down from their original resolution.