Recall Z-1227-2013— Stryker Endoscopy

Class II · Terminated

Class II FDA medical device recall by Stryker Endoscopy, initiated 2013-03-27, terminated 2013-09-24. It names one FDA 510(k) clearance: K954726. Product: Stryker Irrigator Reusable Tips: 2mm X 32cm Regular Tip with 2 Hole, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.. Reason: The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C..

Recalling firm
Stryker Endoscopy
Classification
Class II
Status
Terminated
Initiated
2013-03-27
Terminated
2013-09-24
Firm location
San Jose, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Stryker Irrigator Reusable Tips: 2mm X 32cm Regular Tip with 2 Hole, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.

Reason for recall

The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.

Data sourced from openFDA. This site is unofficial and independent of the FDA.