Recall Z-1226-2014— Ortho-Clinical Diagnostics
Class II · Terminated
Class II FDA medical device recall by Ortho-Clinical Diagnostics, initiated 2014-01-09, terminated 2015-07-07. It names 2 FDA 510(k) clearances: K922072, K840880. Product: VITROS Chemistry Products White Reference Slides, White Correction Factor Slide (WCF), DT Slides, 25 Slides, Part Number J02315; Made in USA by Ortho- Clinical Diagnostics, Inc. 100 Indigo Creek Drive, Rochester, NY 14626 --- This product is a part/ tool used when performing the correction factors adjustment on the VITROS Chemistry Systems. It does not have a 510(k) number. Because the WCF slide is a part/ tool, and is not a device used for diagnostic testing, there is no Intended Use statement. The following summary is provided as a description/ explanation of the WCF slide. The WCF slide is a part/ tool used in the reflectometer correction factor adjustment procedure in order to optimize the optical calibration of the VITROS 250/350/5,1 FS/4600 Chemistry Systems and the VITROS 5600 Integrated. Reason: An incorrect assay value (0.3785) was printed on the product carton for the 460 nm wavelength (filter) that may cause a delay in the ability to obtain and report TBIL, Bu, Bc, and derived tests results. The correct assay value is 0.8658..
- Recalling firm
- Ortho-Clinical Diagnostics
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-01-09
- Terminated
- 2015-07-07
- Firm location
- Rochester, NY, United States
510(k) clearances named in this recall
2 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
VITROS Chemistry Products White Reference Slides, White Correction Factor Slide (WCF), DT Slides, 25 Slides, Part Number J02315; Made in USA by Ortho- Clinical Diagnostics, Inc. 100 Indigo Creek Drive, Rochester, NY 14626 --- This product is a part/ tool used when performing the correction factors adjustment on the VITROS Chemistry Systems. It does not have a 510(k) number. Because the WCF slide is a part/ tool, and is not a device used for diagnostic testing, there is no Intended Use statement. The following summary is provided as a description/ explanation of the WCF slide. The WCF slide is a part/ tool used in the reflectometer correction factor adjustment procedure in order to optimize the optical calibration of the VITROS 250/350/5,1 FS/4600 Chemistry Systems and the VITROS 5600 Integrated
Reason for recall
An incorrect assay value (0.3785) was printed on the product carton for the 460 nm wavelength (filter) that may cause a delay in the ability to obtain and report TBIL, Bu, Bc, and derived tests results. The correct assay value is 0.8658.
Data sourced from openFDA. This site is unofficial and independent of the FDA.