Recall Z-1225-2015— GE Inspection Technologies, LP

Class II · Terminated

Class II FDA medical device recall by GE Inspection Technologies, LP, initiated 2015-01-13, terminated 2015-08-17. FDA's recall record names no specific 510(k) clearance for this event. Product: GE Inspection Technologies x|argos 160D system. Cabinet x-ray system.. Reason: It was discovered that the upper left corner-shaped lead shield in the window in one of the doors installed had lost adhesion and fall out of position, which could lead to gaps in shielding..

Recalling firm
GE Inspection Technologies, LP
Classification
Class II
Status
Terminated
Initiated
2015-01-13
Terminated
2015-08-17
Firm location
Lewistown, PA, United States

Product description

GE Inspection Technologies x|argos 160D system. Cabinet x-ray system.

Reason for recall

It was discovered that the upper left corner-shaped lead shield in the window in one of the doors installed had lost adhesion and fall out of position, which could lead to gaps in shielding.