Recall Z-1225-2014— Philips Medical Systems North America Inc.

Class II · Terminated

Class II FDA medical device recall by Philips Medical Systems North America Inc., initiated 2014-02-06, terminated 2017-06-29. It names one FDA 510(k) clearance: K111693. Product: Philips HeartStart FR3 Automated External Defibrillator (AED), models 861388 and 861389. The HeartStart FR3 is a battery powered automated external defibrillator (AED) designed for use by responders who have been trained in Basic Life Support (BLS), Advanced Life Support (ALS), or another physician-authorized emergency medical response program. The HeartStart FR3 is used to treat suspected victims of ventricular fibrillation (VF), the most common cause of sudden cardiac arrest (SCA), and certain ventricular tachycardia (VTs). Both models are used with disposable defibrillator pads applied to potential victims of SCA with the following symptoms: -Unresponsiveness -Absence of normal breathing If in doubt, apply the pads. The HeartStart FR3 is intended for adults and children over 55 pounds (25 kg) or 8 years old. Both models 861388 and 861389 are also intended for children under 55 pounds (25 kg) or 8 years old when used with the optional Infant/Child Key. If the Infant/Child Key is not available, or you are uncertain of the child's age or weight, do not delay treatment.. Reason: Philip's is recalling HeartStart FR3 Automated External Defibrillator (AED) because the Printed Circuit Board (PCB), which contains the circuitry that supplies energy to the device, could become separated from the electrical contact points that hold the board in place. In the event the PCB becomes separated from the electrical contact points, the AED may not be ready to deliver a shock in the event of an emergency..

Recalling firm
Philips Medical Systems North America Inc.
Classification
Class II
Status
Terminated
Initiated
2014-02-06
Terminated
2017-06-29
Firm location
Bothell, WA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Philips HeartStart FR3 Automated External Defibrillator (AED), models 861388 and 861389. The HeartStart FR3 is a battery powered automated external defibrillator (AED) designed for use by responders who have been trained in Basic Life Support (BLS), Advanced Life Support (ALS), or another physician-authorized emergency medical response program. The HeartStart FR3 is used to treat suspected victims of ventricular fibrillation (VF), the most common cause of sudden cardiac arrest (SCA), and certain ventricular tachycardia (VTs). Both models are used with disposable defibrillator pads applied to potential victims of SCA with the following symptoms: -Unresponsiveness -Absence of normal breathing If in doubt, apply the pads. The HeartStart FR3 is intended for adults and children over 55 pounds (25 kg) or 8 years old. Both models 861388 and 861389 are also intended for children under 55 pounds (25 kg) or 8 years old when used with the optional Infant/Child Key. If the Infant/Child Key is not available, or you are uncertain of the child's age or weight, do not delay treatment.

Reason for recall

Philip's is recalling HeartStart FR3 Automated External Defibrillator (AED) because the Printed Circuit Board (PCB), which contains the circuitry that supplies energy to the device, could become separated from the electrical contact points that hold the board in place. In the event the PCB becomes separated from the electrical contact points, the AED may not be ready to deliver a shock in the event of an emergency.

Data sourced from openFDA. This site is unofficial and independent of the FDA.