Recall Z-1218-2019— Smiths Medical ASD Inc.

Class II · Terminated

Class II FDA medical device recall by Smiths Medical ASD Inc., initiated 2018-08-22, terminated 2021-02-11. It names one FDA 510(k) clearance: K160235. Product: Jelco(R) ViaValve(TM) safety intravenous (I.V.) catheters, 18G x 1 1/4", Model Number REF 3265, MOD 10 The Jelco(R) ViaValve(TM) safety I.V. catheter is a sterile, non pyrogenic, single use, radiopaque intravascular catheter device used for the administration of medically prescribed fluids, designed to reduce the risk of accidental secondary needle sticks.. Reason: Several lots of Jelco¿ ViaValve" safety intravenous (I.V.) catheters exhibited a V shaped puncture at the base of the catheter near the hub due to an adjustment that was made in the manufacturing process..

Recalling firm
Smiths Medical ASD Inc.
Classification
Class II
Status
Terminated
Initiated
2018-08-22
Terminated
2021-02-11
Firm location
Minneapolis, MN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Jelco(R) ViaValve(TM) safety intravenous (I.V.) catheters, 18G x 1 1/4", Model Number REF 3265, MOD 10 The Jelco(R) ViaValve(TM) safety I.V. catheter is a sterile, non pyrogenic, single use, radiopaque intravascular catheter device used for the administration of medically prescribed fluids, designed to reduce the risk of accidental secondary needle sticks.

Reason for recall

Several lots of Jelco¿ ViaValve" safety intravenous (I.V.) catheters exhibited a V shaped puncture at the base of the catheter near the hub due to an adjustment that was made in the manufacturing process.

Data sourced from openFDA. This site is unofficial and independent of the FDA.