Recall Z-1217-2026— CareFusion 303, Inc.
Class II · Ongoing
Class II FDA medical device recall by CareFusion 303, Inc., initiated 2025-11-06. It names one FDA 510(k) clearance: K211218. Product: BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, pediatric, and neonatal patients through clinically accepted routes of administration: intravenous (IV), intra-arterial (IA), subcutaneous, epidural, or irrigation of fluid spaces. The BD AlarisTM System includes the PC Unit (PCU) and one or more of the following: Pump Module, Syringe Module, end-tidal CO2 (EtCO2) Module, Auto-ID Module, patient-controlled analgesia (PCA) Module, and associated software applications.. Reason: Due to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with hospital networks. This issue only affects the Interoperability workflow for devices with a DST invalid timestamp, impacting connectivity of the PCU to the hospital network.
- Recalling firm
- CareFusion 303, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-11-06
- Firm location
- San Diego, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, pediatric, and neonatal patients through clinically accepted routes of administration: intravenous (IV), intra-arterial (IA), subcutaneous, epidural, or irrigation of fluid spaces. The BD AlarisTM System includes the PC Unit (PCU) and one or more of the following: Pump Module, Syringe Module, end-tidal CO2 (EtCO2) Module, Auto-ID Module, patient-controlled analgesia (PCA) Module, and associated software applications.
Reason for recall
Due to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with hospital networks. This issue only affects the Interoperability workflow for devices with a DST invalid timestamp, impacting connectivity of the PCU to the hospital network
Data sourced from openFDA. This site is unofficial and independent of the FDA.