Recall Z-1217-2014— Exactech, Inc.
Class II · Terminated
Class II FDA medical device recall by Exactech, Inc., initiated 2014-01-06, terminated 2014-08-04. It names one FDA 510(k) clearance: K110197. Product: Gibralt Spine System Polyaxial Screw Intended to promote fusion of the cervical spine, and the thoracic spine, (C3-T3), the Gibralt Spinal System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors. The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C3-T3) spine.. Reason: The products have been identified to contain springs which have been manufactured with type-304 stainless steel rather than the specified titanium alloy..
- Recalling firm
- Exactech, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-01-06
- Terminated
- 2014-08-04
- Firm location
- Gainesville, FL, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Gibralt Spine System Polyaxial Screw Intended to promote fusion of the cervical spine, and the thoracic spine, (C3-T3), the Gibralt Spinal System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors. The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C3-T3) spine.
Reason for recall
The products have been identified to contain springs which have been manufactured with type-304 stainless steel rather than the specified titanium alloy.