Recall Z-1213-2019— COVIDIEN LLC
Class II · Terminated
Class II FDA medical device recall by COVIDIEN LLC, initiated 2019-03-13, terminated 2020-04-23. It names one FDA 510(k) clearance: K903205. Product: Procedure kits include devices from the affected lots distributed OUS only: (1)BOX KIT00529H KIT COLECISTECTOMIA ACCESO (2)BOX KIT00680 KIT COLECISTECTOMIA- (3)BOX KIT00681 KIT APENDICECTOMIA (4)BOX KIT00897 KIT COLON IZQUIERDO (5)BOX LAPL1 ZESTAW DO CHOLECYSTEKTOMII (6)BOX PST01134 LAP CHOLE KIT ST JOHN S (7)KIT00599 LAPARO GENERICO X1 (8)KIT00608 APENDICECTOMIA LAP X1 (9)KIT2176CV BORNHOLM LAP INGUINAL HERNIA (10)KIT2951V CHOLECYSTECT BORNHOLM (11KIT2963 APPENDECTOMY KIT X1 (12)KITM025 M'LANDS MR NASSAR LAP CHOLEX1 The Covidien Endo Grasp" Auto Suture" grasper 5 mm instrument has applications in a variety of gynecologic, general, urologic, thoracic and endoscopic procedures for temporary grasping/clamping of tissue and small tubular structures. 1)BOX KIT00529H KIT COLECISTECTOMIA ACCESO Lot Number: 0217118756 (2)BOX KIT00680 KIT COLECISTECTOMIA- 0217117865 0217118746 0217117863 0217117858 0217117864 0217117862 0217117861 0217117859 0217117857 0217117860 (3)BOX KIT00681 KIT APENDICECTOMIA 0217118750 0217118749 0217118748 0217118747 (4)BOX KIT00897 KIT COLON IZQUIERDO 0217134109 (5)BOX LAPL1 ZESTAW DO CHOLECYSTEKTOMII 0217118761 (6)BOX PST01134 LAP CHOLE KIT ST JOHN S 0217119261, 0217119259 (7)KIT00599 LAPARO GENERICO X1 0217130753, 0217130754 (8)KIT00608 APENDICECTOMIA LAP X1 0217132887 (9)KIT2176CV BORNHOLM LAP INGUINAL HERNIA 0217123007 (10)KIT2951V CHOLECYSTECT BORNHOLM 0217123010 (11KIT2963 APPENDECTOMY KIT X1 0217123009 (12)KITM025 M'LANDS MR NASSAR LAP CHOLEX1 0217127223 (1)BOX KIT00529H KIT COLECISTECTOMIA ACCESO Lot Number: 0217118756 (2)BOX KIT00680 KIT COLECISTECTOMIA- 0217117865 0217118746 0217117863 0217117858 0217117864 0217117862 0217117861 0217117859 0217117857 0217117860 (3)BOX KIT00681 KIT APENDICECTOMIA 0217118750 0217118749 0217118748 0217118747 (4)BOX KIT00897 KIT COLON IZQUIERDO 0217134109 (5)BOX LAPL1 ZESTAW DO CHOLECYSTEKTOMII 0217118761 (6)BOX PST01134 LAP CHOLE KIT ST JOHN S 0217119261, 0217119259 (7)KIT00599 LAPARO GENERICO X1 0217130753, 0217130754 (8)KIT00608 APENDICECTOMIA LAP X1 0217132887 (9)KIT2176CV BORNHOLM LAP INGUINAL HERNIA 0217123007 (10)KIT2951V CHOLECYSTECT BORNHOLM 0217123010 (11KIT2963 APPENDECTOMY KIT X1 0217123009 (12)KITM025 M'LANDS MR NASSAR LAP CHOLEX1 0217127223 (1)BOX KIT00529H KIT COLECISTECTOMIA ACCESO Lot Number: 0217118756 (2)BOX KIT00680 KIT COLECISTECTOMIA- 0217117865 0217118746 0217117863 0217117858 0217117864 0217117862 0217117861 0217117859 0217117857 0217117860 (3)BOX KIT00681 KIT APENDICECTOMIA 0217118750 0217118749 0217118748 0217118747 (4)BOX KIT00897 KIT COLON IZQUIERDO 0217134109 (5)BOX LAPL1 ZESTAW DO CHOLECYSTEKTOMII 0217118761 (6)BOX PST01134 LAP CHOLE KIT ST JOHN S 0217119261, 0217119259 (7)KIT00599 LAPARO GENERICO X1 0217130753, 0217130754 (8)KIT00608 APENDICECTOMIA LAP X1 0217132887 (9)KIT2176CV BORNHOLM LAP INGUINAL HERNIA 0217123007 (10)KIT2951V CHOLECYSTECT BORNHOLM 0217123010 (11KIT2963 APPENDECTOMY KIT X1 0217123009 (12)KITM025 M'LANDS MR NASSAR LAP CHOLEX1 0217127223. Reason: Sterilization method used was not consistent with the labeling and the approved sterilization method.
- Recalling firm
- COVIDIEN LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-03-13
- Terminated
- 2020-04-23
- Firm location
- Mansfield, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Procedure kits include devices from the affected lots distributed OUS only: (1)BOX KIT00529H KIT COLECISTECTOMIA ACCESO (2)BOX KIT00680 KIT COLECISTECTOMIA- (3)BOX KIT00681 KIT APENDICECTOMIA (4)BOX KIT00897 KIT COLON IZQUIERDO (5)BOX LAPL1 ZESTAW DO CHOLECYSTEKTOMII (6)BOX PST01134 LAP CHOLE KIT ST JOHN S (7)KIT00599 LAPARO GENERICO X1 (8)KIT00608 APENDICECTOMIA LAP X1 (9)KIT2176CV BORNHOLM LAP INGUINAL HERNIA (10)KIT2951V CHOLECYSTECT BORNHOLM (11KIT2963 APPENDECTOMY KIT X1 (12)KITM025 M'LANDS MR NASSAR LAP CHOLEX1 The Covidien Endo Grasp" Auto Suture" grasper 5 mm instrument has applications in a variety of gynecologic, general, urologic, thoracic and endoscopic procedures for temporary grasping/clamping of tissue and small tubular structures. 1)BOX KIT00529H KIT COLECISTECTOMIA ACCESO Lot Number: 0217118756 (2)BOX KIT00680 KIT COLECISTECTOMIA- 0217117865 0217118746 0217117863 0217117858 0217117864 0217117862 0217117861 0217117859 0217117857 0217117860 (3)BOX KIT00681 KIT APENDICECTOMIA 0217118750 0217118749 0217118748 0217118747 (4)BOX KIT00897 KIT COLON IZQUIERDO 0217134109 (5)BOX LAPL1 ZESTAW DO CHOLECYSTEKTOMII 0217118761 (6)BOX PST01134 LAP CHOLE KIT ST JOHN S 0217119261, 0217119259 (7)KIT00599 LAPARO GENERICO X1 0217130753, 0217130754 (8)KIT00608 APENDICECTOMIA LAP X1 0217132887 (9)KIT2176CV BORNHOLM LAP INGUINAL HERNIA 0217123007 (10)KIT2951V CHOLECYSTECT BORNHOLM 0217123010 (11KIT2963 APPENDECTOMY KIT X1 0217123009 (12)KITM025 M'LANDS MR NASSAR LAP CHOLEX1 0217127223 (1)BOX KIT00529H KIT COLECISTECTOMIA ACCESO Lot Number: 0217118756 (2)BOX KIT00680 KIT COLECISTECTOMIA- 0217117865 0217118746 0217117863 0217117858 0217117864 0217117862 0217117861 0217117859 0217117857 0217117860 (3)BOX KIT00681 KIT APENDICECTOMIA 0217118750 0217118749 0217118748 0217118747 (4)BOX KIT00897 KIT COLON IZQUIERDO 0217134109 (5)BOX LAPL1 ZESTAW DO CHOLECYSTEKTOMII 0217118761 (6)BOX PST01134 LAP CHOLE KIT ST JOHN S 0217119261, 0217119259 (7)KIT00599 LAPARO GENERICO X1 0217130753, 0217130754 (8)KIT00608 APENDICECTOMIA LAP X1 0217132887 (9)KIT2176CV BORNHOLM LAP INGUINAL HERNIA 0217123007 (10)KIT2951V CHOLECYSTECT BORNHOLM 0217123010 (11KIT2963 APPENDECTOMY KIT X1 0217123009 (12)KITM025 M'LANDS MR NASSAR LAP CHOLEX1 0217127223 (1)BOX KIT00529H KIT COLECISTECTOMIA ACCESO Lot Number: 0217118756 (2)BOX KIT00680 KIT COLECISTECTOMIA- 0217117865 0217118746 0217117863 0217117858 0217117864 0217117862 0217117861 0217117859 0217117857 0217117860 (3)BOX KIT00681 KIT APENDICECTOMIA 0217118750 0217118749 0217118748 0217118747 (4)BOX KIT00897 KIT COLON IZQUIERDO 0217134109 (5)BOX LAPL1 ZESTAW DO CHOLECYSTEKTOMII 0217118761 (6)BOX PST01134 LAP CHOLE KIT ST JOHN S 0217119261, 0217119259 (7)KIT00599 LAPARO GENERICO X1 0217130753, 0217130754 (8)KIT00608 APENDICECTOMIA LAP X1 0217132887 (9)KIT2176CV BORNHOLM LAP INGUINAL HERNIA 0217123007 (10)KIT2951V CHOLECYSTECT BORNHOLM 0217123010 (11KIT2963 APPENDECTOMY KIT X1 0217123009 (12)KITM025 M'LANDS MR NASSAR LAP CHOLEX1 0217127223
Reason for recall
Sterilization method used was not consistent with the labeling and the approved sterilization method