Recall Z-1210-2024— Olympus Corporation of the Americas

Class II · Ongoing

Class II FDA medical device recall by Olympus Corporation of the Americas, initiated 2024-01-11. It names one FDA 510(k) clearance: K131780. Product: Colonoscope, Model Number PCF-H190DL.. Reason: Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads..

Recalling firm
Olympus Corporation of the Americas
Classification
Class II
Status
Ongoing
Initiated
2024-01-11
Firm location
Center Valley, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Colonoscope, Model Number PCF-H190DL.

Reason for recall

Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.

Data sourced from openFDA. This site is unofficial and independent of the FDA.