Recall Z-1206-2013— Cardinal Health, Medical Products & Services

Class I · Terminated

Class I FDA medical device recall by Cardinal Health, Medical Products & Services, initiated 2013-03-26, terminated 2017-04-12. FDA's recall record names no specific 510(k) clearance for this event. Product: Presource PBDS, Knee Arthroscopy, Kit, Circulator. Reason: Various Presource Kits contain a pre-assembled Filter and Anesthesia Circuit that may contain outer plastic packaging on one or more components. If the packaging is removed without disassembling the components, remnants of the plastic from the packaging material may become lodged in the filter potentially causing an obstruction of airflow..

Recalling firm
Cardinal Health, Medical Products & Services
Classification
Class I
Status
Terminated
Initiated
2013-03-26
Terminated
2017-04-12
Firm location
Park City, IL, United States

Product description

Presource PBDS, Knee Arthroscopy, Kit, Circulator

Reason for recall

Various Presource Kits contain a pre-assembled Filter and Anesthesia Circuit that may contain outer plastic packaging on one or more components. If the packaging is removed without disassembling the components, remnants of the plastic from the packaging material may become lodged in the filter potentially causing an obstruction of airflow.

Data sourced from openFDA. This site is unofficial and independent of the FDA.