Recall Z-1205-2019— Brainlab AG

Class II · Terminated

Class II FDA medical device recall by Brainlab AG, initiated 2019-03-11, terminated 2021-05-12. It names 2 FDA 510(k) clearances: K170750, K142108. Product: RT Elements Software revisions of the RT Elements applications have a specific software version number. Specifically the following RT Elements applications/versions are affected: - Cranial SRS 1.0.0 and 1.5.0 - Spine SRS 1.0.0 and 1.5.0 - Multiple Brain Mets SRS 1.5.0 - RT QA 1.0.0 and 1.5.0 Product Usage: The RT Elements are applications for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck and extracranial lesions.. Reason: There is a potential for an incorrect dose distribution calculation by Brainlab RT Elements software (for affected versions) under specific circumstances when using the Pencil Beam algorithm on the GPU (graphics card), as is the default system setting..

Recalling firm
Brainlab AG
Classification
Class II
Status
Terminated
Initiated
2019-03-11
Terminated
2021-05-12
Firm location
Munich, N/A, Germany

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

RT Elements Software revisions of the RT Elements applications have a specific software version number. Specifically the following RT Elements applications/versions are affected: - Cranial SRS 1.0.0 and 1.5.0 - Spine SRS 1.0.0 and 1.5.0 - Multiple Brain Mets SRS 1.5.0 - RT QA 1.0.0 and 1.5.0 Product Usage: The RT Elements are applications for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck and extracranial lesions.

Reason for recall

There is a potential for an incorrect dose distribution calculation by Brainlab RT Elements software (for affected versions) under specific circumstances when using the Pencil Beam algorithm on the GPU (graphics card), as is the default system setting.