Recall Z-1203-2026— Cardinal Health 200, LLC
Class II · Ongoing
Class II FDA medical device recall by Cardinal Health 200, LLC, initiated 2025-12-26. FDA's recall record names no specific 510(k) clearance for this event. Product: Chest Drainage Units and Accessories: Product Code Description 1. 8888571562 Cardinal Health Sentinel Seal CDU 2. 8888571513 Cardinal Health Sentinel Seal CDU, Dual Drain 3. 8888571489 Cardinal Health Sentinel Seal CDU, with Easy Change Connector 4. 8888571370 Cardinal Health Altitude CDU, Dry Suction, with Easy Change Connector 5. 8888571299 Cardinal Health Aqua-Seal CDU, Wet Suction 6. 8888571406 Cardinal Health Aqua-Seal CDU, Wet Suction, Dual Drain 7. 8888571315 Cardinal Health Aqua-Seal CDU, Wet Suction, with Easy Change Connector 8. 8884713100 Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL 9. 1814713105 OUS¿Only - Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL (Bottle only) 10. 1180571570 OUS¿Only - Cardinal Health Thora-Seal CDU, Two Chamber System, 2600 mL 11. 8884713308 Cardinal Health Thora-Seal CDU, Three Chamber System, 2500 mL 12. 8884713900 Cardinal Health Thora-Seal Replacement Collection Chamber, 2500 mL. Reason: The firm is updating the IFU to clarify the intended patient population for the identified chest drainage units and accessories as adults, 18 years and older. When used on infants, the CDU may not demonstrate tidaling/bubbling, which may confound the clinician's assessment of device functionality and lead to inadequate/inappropriate treatment/therapy, delay to treatment/therapy, and prolonged hospitalization..
- Recalling firm
- Cardinal Health 200, LLC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-12-26
- Firm location
- Waukegan, IL, United States
Product description
Chest Drainage Units and Accessories: Product Code Description 1. 8888571562 Cardinal Health Sentinel Seal CDU 2. 8888571513 Cardinal Health Sentinel Seal CDU, Dual Drain 3. 8888571489 Cardinal Health Sentinel Seal CDU, with Easy Change Connector 4. 8888571370 Cardinal Health Altitude CDU, Dry Suction, with Easy Change Connector 5. 8888571299 Cardinal Health Aqua-Seal CDU, Wet Suction 6. 8888571406 Cardinal Health Aqua-Seal CDU, Wet Suction, Dual Drain 7. 8888571315 Cardinal Health Aqua-Seal CDU, Wet Suction, with Easy Change Connector 8. 8884713100 Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL 9. 1814713105 OUS¿Only - Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL (Bottle only) 10. 1180571570 OUS¿Only - Cardinal Health Thora-Seal CDU, Two Chamber System, 2600 mL 11. 8884713308 Cardinal Health Thora-Seal CDU, Three Chamber System, 2500 mL 12. 8884713900 Cardinal Health Thora-Seal Replacement Collection Chamber, 2500 mL
Reason for recall
The firm is updating the IFU to clarify the intended patient population for the identified chest drainage units and accessories as adults, 18 years and older. When used on infants, the CDU may not demonstrate tidaling/bubbling, which may confound the clinician's assessment of device functionality and lead to inadequate/inappropriate treatment/therapy, delay to treatment/therapy, and prolonged hospitalization.
Data sourced from openFDA. This site is unofficial and independent of the FDA.