Recall Z-1201-2026— Zimmer Surgical Inc
Class II · Ongoing
Class II FDA medical device recall by Zimmer Surgical Inc, initiated 2025-12-24. It names one FDA 510(k) clearance: K202919. Product: Brand Name: Zimmer Tourniquet Systems Product Name: A.T.S.¿ 5000TS Tourniquet Systems Model/Catalog Number: 60500010100 Software Version: Software version v2.04 or prior are within scope of the recall for A.T.S.¿ 5000TS Product Description: A.T.S 5000TS Tourniquet Systems Component: N/A. Reason: Device user interface (screen) may freeze or become unresponsive to touch when operating in certain non-English language settings. Issue may cause delay or extension while alternate device is located..
- Recalling firm
- Zimmer Surgical Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-12-24
- Firm location
- Dover, OH, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Brand Name: Zimmer Tourniquet Systems Product Name: A.T.S.¿ 5000TS Tourniquet Systems Model/Catalog Number: 60500010100 Software Version: Software version v2.04 or prior are within scope of the recall for A.T.S.¿ 5000TS Product Description: A.T.S 5000TS Tourniquet Systems Component: N/A
Reason for recall
Device user interface (screen) may freeze or become unresponsive to touch when operating in certain non-English language settings. Issue may cause delay or extension while alternate device is located.
Data sourced from openFDA. This site is unofficial and independent of the FDA.