Recall Z-1199-2017— Acumed LLC

Class II · Terminated

Class II FDA medical device recall by Acumed LLC, initiated 2017-01-19, terminated 2017-05-08. It names one FDA 510(k) clearance: K122538. Product: 6-hole Pubic Symphysis plate (PN 70-0451). The pelvic bone plates for the pelvic ring are a series of plates with varying lengths that function as internal fixation devices for fractures, fusions and osteotomies of the pelvic ring. The plates are strategically pre-contoured where beneficial to the application, and they are secured to the bone with 3.5mm non-locking screws. The Acumed pelvic ring plate is the Pubic Symphysis Plate. The Acumed Pelvic Plating System is intended for use by surgeons with orthopedic training and knowledge of the indications and techniques required for fixation. The device is to be implanted by the surgeon in a sterile operating room setting.. Reason: The 6-hole Pubic Symphysis plate (PN 70-0451) is specified to be made of Commercially Pure Titanium (ASTM F67 - CP Grade 4). However, batch 383813 was manufactured out of Titanium Alloy (Ti-6Al- 4V ELI per ASTM F135  Ti alloy) and was distributed..

Recalling firm
Acumed LLC
Classification
Class II
Status
Terminated
Initiated
2017-01-19
Terminated
2017-05-08
Firm location
Hillsboro, OR, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

6-hole Pubic Symphysis plate (PN 70-0451). The pelvic bone plates for the pelvic ring are a series of plates with varying lengths that function as internal fixation devices for fractures, fusions and osteotomies of the pelvic ring. The plates are strategically pre-contoured where beneficial to the application, and they are secured to the bone with 3.5mm non-locking screws. The Acumed pelvic ring plate is the Pubic Symphysis Plate. The Acumed Pelvic Plating System is intended for use by surgeons with orthopedic training and knowledge of the indications and techniques required for fixation. The device is to be implanted by the surgeon in a sterile operating room setting.

Reason for recall

The 6-hole Pubic Symphysis plate (PN 70-0451) is specified to be made of Commercially Pure Titanium (ASTM F67 - CP Grade 4). However, batch 383813 was manufactured out of Titanium Alloy (Ti-6Al- 4V ELI per ASTM F135  Ti alloy) and was distributed.