Recall Z-1197-2015— Navilyst Medical, Inc

Class II · Terminated

Class II FDA medical device recall by Navilyst Medical, Inc, initiated 2015-01-09, terminated 2017-03-10. FDA's recall record names no specific 510(k) clearance for this event. Product: NAMIC Convenience Kit, 3V MANIFOLD KIT, UPN H965601011621, REF/Catalog No. 60101162, STERILE, Rx ONLY ---For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.. Reason: The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection..

Recalling firm
Navilyst Medical, Inc
Classification
Class II
Status
Terminated
Initiated
2015-01-09
Terminated
2017-03-10
Firm location
Glens Falls, NY, United States

Product description

NAMIC Convenience Kit, 3V MANIFOLD KIT, UPN H965601011621, REF/Catalog No. 60101162, STERILE, Rx ONLY ---For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.

Reason for recall

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Data sourced from openFDA. This site is unofficial and independent of the FDA.