Recall Z-1195-2025— RAYSEARCH LABORATORIES AB

Class II · Ongoing

Class II FDA medical device recall by RAYSEARCH LABORATORIES AB, initiated 2025-01-16. It names one FDA 510(k) clearance: K141860. Product: Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 4.5.1.14, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4 and 4.7.6.7 Software Version: RayStation 4.5.1, 4.7.2, 4.7.3, 4.7.4, 4.7.5 and 4.7.6. Reason: Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs..

Recalling firm
RAYSEARCH LABORATORIES AB
Classification
Class II
Status
Ongoing
Initiated
2025-01-16
Firm location
Stockholm, N/A, Sweden

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 4.5.1.14, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4 and 4.7.6.7 Software Version: RayStation 4.5.1, 4.7.2, 4.7.3, 4.7.4, 4.7.5 and 4.7.6

Reason for recall

Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.

Data sourced from openFDA. This site is unofficial and independent of the FDA.