Recall Z-1193-2025— RAYSEARCH LABORATORIES AB

Class II · Ongoing

Class II FDA medical device recall by RAYSEARCH LABORATORIES AB, initiated 2025-01-16. It names one FDA 510(k) clearance: K170355. Product: Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 6.0.0.24 Software Version: RayStation 6. Reason: Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs..

Recalling firm
RAYSEARCH LABORATORIES AB
Classification
Class II
Status
Ongoing
Initiated
2025-01-16
Firm location
Stockholm, N/A, Sweden

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 6.0.0.24 Software Version: RayStation 6

Reason for recall

Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.

Data sourced from openFDA. This site is unofficial and independent of the FDA.