Recall Z-1191-2023— Stryker GmbH
Class II · Ongoing
Class II FDA medical device recall by Stryker GmbH, initiated 2023-02-07. It names one FDA 510(k) clearance: K153377. Product: stryker Hoffmann LRF, Transport Strut, Model Number 4933-0-400; Orthopedic external fixation device intended to keep bone fragments/segments in a correct anatomical position until reliable bone consolidation is achieved. Reason: Stryker has identified a nonconformance in specific lots of Hoffmann LRF Bone Transport Struts. Specifically, the thread pitch is oversized at the distal end of the threaded rod, and as a result the distal ring fixation cannot be locked between 0- 50mm of the threaded rod..
- Recalling firm
- Stryker GmbH
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-02-07
- Firm location
- Selzach, N/A, Switzerland
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
stryker Hoffmann LRF, Transport Strut, Model Number 4933-0-400; Orthopedic external fixation device intended to keep bone fragments/segments in a correct anatomical position until reliable bone consolidation is achieved
Reason for recall
Stryker has identified a nonconformance in specific lots of Hoffmann LRF Bone Transport Struts. Specifically, the thread pitch is oversized at the distal end of the threaded rod, and as a result the distal ring fixation cannot be locked between 0- 50mm of the threaded rod.