Recall Z-1191-2017— Encore Medical, Lp
Class II · Terminated
Class II FDA medical device recall by Encore Medical, Lp, initiated 2017-01-10, terminated 2017-03-31. It names one FDA 510(k) clearance: K092331. Product: LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up of a modular stem coupled with a proper neck by means of a Morse taper stabilized during the implantation phase by a safety screw. This system is particularly indicated for revision surgery on both uncemented and cemented femoral implants, when there is significant bone loss and an abnormal meta-epiphyseal anatomy of the femur.. Reason: Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma sterilization..
- Recalling firm
- Encore Medical, Lp
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-01-10
- Terminated
- 2017-03-31
- Firm location
- Austin, TX, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up of a modular stem coupled with a proper neck by means of a Morse taper stabilized during the implantation phase by a safety screw. This system is particularly indicated for revision surgery on both uncemented and cemented femoral implants, when there is significant bone loss and an abnormal meta-epiphyseal anatomy of the femur.
Reason for recall
Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma sterilization.
Data sourced from openFDA. This site is unofficial and independent of the FDA.