Recall Z-1188-2025— RAYSEARCH LABORATORIES AB
Class II · Ongoing
Class II FDA medical device recall by RAYSEARCH LABORATORIES AB, initiated 2025-01-16. It names one FDA 510(k) clearance: K210645. Product: Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 10.1.0.613 and 10.1.1.54 Software Version: RayStation 10B and 10B Service Pack 1. Reason: Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs..
- Recalling firm
- RAYSEARCH LABORATORIES AB
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-01-16
- Firm location
- Stockholm, N/A, Sweden
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 10.1.0.613 and 10.1.1.54 Software Version: RayStation 10B and 10B Service Pack 1
Reason for recall
Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
Data sourced from openFDA. This site is unofficial and independent of the FDA.