Recall Z-1187-2013— Masimo Corporation

Class II · Terminated

Class II FDA medical device recall by Masimo Corporation, initiated 2013-03-13, terminated 2015-06-10. FDA's recall record names no specific 510(k) clearance for this event. Product: Rad-8 Pulse Oximeters The Rad-8 family of pulse oximeters and accessories are indicated for the continuous, noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor).. Reason: Masimo has identified a very small number (0.02%) of Rad-8 devices with 20-pin patient cable connector that, when moved, can power off without the operator pressing the power button..

Recalling firm
Masimo Corporation
Classification
Class II
Status
Terminated
Initiated
2013-03-13
Terminated
2015-06-10
Firm location
Irvine, CA, United States

Product description

Rad-8 Pulse Oximeters The Rad-8 family of pulse oximeters and accessories are indicated for the continuous, noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor).

Reason for recall

Masimo has identified a very small number (0.02%) of Rad-8 devices with 20-pin patient cable connector that, when moved, can power off without the operator pressing the power button.