Recall Z-1185-2023— Stryker Corporation

Class II · Ongoing

Class II FDA medical device recall by Stryker Corporation, initiated 2023-01-25. It names one FDA 510(k) clearance: K142801. Product: Finished Goods Box of 50 Inserts and Scanning Labels for Medivac Guardian 3L Catalog Number: FG 12022. Intended adjunct in the estimation of blood loss.. Reason: Triton Canister Software, Insert & Scanning Label may cause the Triton Canister System to output inaccurate blood loss estimates can lead to confusion and could potentially result in either delayed recognition or prolongation of postpartum hemorrhage (PPH), or require more aggressive treatment for PPH..

Recalling firm
Stryker Corporation
Classification
Class II
Status
Ongoing
Initiated
2023-01-25
Firm location
Portage, MI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Finished Goods Box of 50 Inserts and Scanning Labels for Medivac Guardian 3L Catalog Number: FG 12022. Intended adjunct in the estimation of blood loss.

Reason for recall

Triton Canister Software, Insert & Scanning Label may cause the Triton Canister System to output inaccurate blood loss estimates can lead to confusion and could potentially result in either delayed recognition or prolongation of postpartum hemorrhage (PPH), or require more aggressive treatment for PPH.