Recall Z-1184-2013— Covidien LLC

Class II · Terminated

Class II FDA medical device recall by Covidien LLC, initiated 2013-04-08, terminated 2015-11-19. It names one FDA 510(k) clearance: K012218. Product: Medi-Trace Cadence Adult Radiotransparent Defibrillation Electrode Product ID: 22550R. Intended for use in cardioversion, pacing, ECG monitoring, and defibrillation procedures.. Reason: Defibrillation Electrodes may arc/spark result in thermal damage to the leadwire, which could render the device incapable of delivering the appropriate energy or shock to the patient.

Recalling firm
Covidien LLC
Classification
Class II
Status
Terminated
Initiated
2013-04-08
Terminated
2015-11-19
Firm location
Mansfield, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Medi-Trace Cadence Adult Radiotransparent Defibrillation Electrode Product ID: 22550R. Intended for use in cardioversion, pacing, ECG monitoring, and defibrillation procedures.

Reason for recall

Defibrillation Electrodes may arc/spark result in thermal damage to the leadwire, which could render the device incapable of delivering the appropriate energy or shock to the patient