Recall Z-1178-2024— PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Class II · Ongoing

Class II FDA medical device recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., initiated 2024-02-06. It names one FDA 510(k) clearance: K200917. Product: Azurion 7 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722223, (2) 722224, (3) 722225, (4) 722226. Reason: Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure..

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Classification
Class II
Status
Ongoing
Initiated
2024-02-06
Firm location
Best, N/A, Netherlands

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Azurion 7 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722223, (2) 722224, (3) 722225, (4) 722226

Reason for recall

Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.