Recall Z-1175-2024— Abiomed, Inc.
Class II · Terminated
Class II FDA medical device recall by Abiomed, Inc., initiated 2021-03-01, terminated 2025-03-06. FDA's recall record names no specific 510(k) clearance for this event. Product: Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems (cardiovascular). Reason: Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC..
- Recalling firm
- Abiomed, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2021-03-01
- Terminated
- 2025-03-06
- Firm location
- Danvers, MA, United States
Product description
Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems (cardiovascular)
Reason for recall
Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.