Recall Z-1175-2016— Toshiba American Medical Systems Inc

Class II · Terminated

Class II FDA medical device recall by Toshiba American Medical Systems Inc, initiated 2016-01-14, terminated 2017-02-24. It names one FDA 510(k) clearance: K133553. Product: Toshiba DRAD-3000E FPD Wireless System Product Usage The DRAD-3000¿ is intended for use with the ceiling-suspended tube support, high voltage generator, and bucky stand or bucky table incorporating a fixed or detachable (portable) flat panel detector for radiography of the head, chest, abdomen, spine, neck, and limbs. This system is used for image acquisition, image display and transmission/output or images to external devices.. Reason: When a user performed radiography using the wireless FPD, a message window appeared on the monitor stating that imaging transmission was not completed with no radiographic image. It also indicated to select the "OK" button to re-acquire the image data or to select the¿'Cancel" button to cancel the re-acquisition. As instructed the user selected "OK" and the same message window appeared. This same operation was repeated several times with the identical result - no image. Finally, the user selected "Cancel" and the re-acquisition was terminated..

Recalling firm
Toshiba American Medical Systems Inc
Classification
Class II
Status
Terminated
Initiated
2016-01-14
Terminated
2017-02-24
Firm location
Tustin, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Toshiba DRAD-3000E FPD Wireless System Product Usage The DRAD-3000¿ is intended for use with the ceiling-suspended tube support, high voltage generator, and bucky stand or bucky table incorporating a fixed or detachable (portable) flat panel detector for radiography of the head, chest, abdomen, spine, neck, and limbs. This system is used for image acquisition, image display and transmission/output or images to external devices.

Reason for recall

When a user performed radiography using the wireless FPD, a message window appeared on the monitor stating that imaging transmission was not completed with no radiographic image. It also indicated to select the "OK" button to re-acquire the image data or to select the¿'Cancel" button to cancel the re-acquisition. As instructed the user selected "OK" and the same message window appeared. This same operation was repeated several times with the identical result - no image. Finally, the user selected "Cancel" and the re-acquisition was terminated.

Data sourced from openFDA. This site is unofficial and independent of the FDA.