Recall Z-1174-2026— Siemens Medical Solutions USA, Inc
Class II · Ongoing
Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2025-12-29. It names one FDA 510(k) clearance: K250660. Product: Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix R; Siemens Material Number (SMN): 11574002; With Software Version VB10D-SP02. Reason: Ortho images acquired with preset automatic horizontal flip are not flipped and when acquiring single images, there is the possibility that an image of a previous patient may be processed..
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-12-29
- Firm location
- Malvern, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix R; Siemens Material Number (SMN): 11574002; With Software Version VB10D-SP02
Reason for recall
Ortho images acquired with preset automatic horizontal flip are not flipped and when acquiring single images, there is the possibility that an image of a previous patient may be processed.
Data sourced from openFDA. This site is unofficial and independent of the FDA.