Recall Z-1173-2018— Biomet, Inc.
Class II · Terminated
Class II FDA medical device recall by Biomet, Inc., initiated 2018-02-28, terminated 2020-08-11. It names one FDA 510(k) clearance: K111746. Product: Zimmer Biomet Comprehensive Segmental Revision System (SRS). Orthopedic implants for Proximal Humeral Reconstruction, Distal Humeral Reconstruction, and Total Humeral Reconstruction. Reason: During assembly, an SRS Humeral Distal Body implant failed to engage correctly at the point of reduction with a Discovery Elbow System Humeral Condyle. A mismatch between the screw holes in the Humeral Body & Elbow Condyle affects assembly during surgery. If the product has already been implanted, no action is required. The mismatch is estimated to prevent screw engagement in less than 1% of cases..
- Recalling firm
- Biomet, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-02-28
- Terminated
- 2020-08-11
- Firm location
- Warsaw, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Zimmer Biomet Comprehensive Segmental Revision System (SRS). Orthopedic implants for Proximal Humeral Reconstruction, Distal Humeral Reconstruction, and Total Humeral Reconstruction
Reason for recall
During assembly, an SRS Humeral Distal Body implant failed to engage correctly at the point of reduction with a Discovery Elbow System Humeral Condyle. A mismatch between the screw holes in the Humeral Body & Elbow Condyle affects assembly during surgery. If the product has already been implanted, no action is required. The mismatch is estimated to prevent screw engagement in less than 1% of cases.
Data sourced from openFDA. This site is unofficial and independent of the FDA.