Recall Z-1171-2014— Biomet 3i, LLC

Class II · Terminated

Class II FDA medical device recall by Biomet 3i, LLC, initiated 2013-07-22, terminated 2015-04-21. FDA's recall record names no specific 510(k) clearance for this event. Product: Contra-Angle Torque Driver Kit For Certain and External Connection Product Usage: Molded plastic tray used in holding various dental instruments for autoclave processing.. Reason: During recent revised sterilization validation testing, the included surgical trays did not meet the Sterility Assurance Level required..

Recalling firm
Biomet 3i, LLC
Classification
Class II
Status
Terminated
Initiated
2013-07-22
Terminated
2015-04-21
Firm location
Palm Beach Gardens, FL, United States

Product description

Contra-Angle Torque Driver Kit For Certain and External Connection Product Usage: Molded plastic tray used in holding various dental instruments for autoclave processing.

Reason for recall

During recent revised sterilization validation testing, the included surgical trays did not meet the Sterility Assurance Level required.

Data sourced from openFDA. This site is unofficial and independent of the FDA.